Yokogawa DR240 User Manual

Page 48

Advertising
background image

3-3

IM DR232-01E

3

Installation and Wiring

• Rack mounting

The following fittings can be supplied for rack mounting.

Name

Model

Description

Rack mount fitting

DV400-013

Conforming to ANSI/EIA

• External dimensions and rack mounting dimensions for the DR232 are shown below.
• To install the rack mount fitting on the instrument, remove the screws at the right and left

forward of the instrument and use the mounting screws attached to the rack mount fitting.
Be careful that right and left screws have different lengths. The screw at the right when
facing the front of the instrument is M4 of 20 mm long and the left screw is M4 of 16 mm
long. Appropriate tightening torque is 1.4 to 1.5 N·m (14 to 15 kgf·cm).

• If another instrument is to be mounted under this instrument in an ANSI/EIA rack, remove

the prong or brackets designated with 2 and 3 in the figure below because they may interfere
with both instruments.

30

207

63.6

1

1

2

438

266

25

Rack Mounting Dementions

11.3

482.6

±

1

265.9

±

1

190.5

±

0.5

37.7

37.7

6.8

6.8

30

Panel face

1&2 : Removed

11.3

265.9

±

1

190.5

±

0.5

3.2 How to Install