12-2

IM DR232-01E

12.1 Overview of the Computation Function

Statistical operators within the group

Type

Operator

Example

Description

Maximum value CLOG.MAX() CLOG.MAX(G01) Obtain the maximum value of the measured data of group G01.

Minimum value CLOG.MIN()

CLOG.MIN(G02) Obtain the minimum value of the measured data of group G02.

Max-min value

CLOG.P-P()

CLOG.P-P(G03)

Obtain the P-P value of the measured data of group G03.

Total value

CLOG.SUM() CLOG.SUM(G04) Obtain the total value of the measured data of group G04.

Average value

CLOG.AVE()

CLOG.AVE(G05) Obtain the average value of the measured data of group G05.

* Statistical computation of the measured data of the input channel within the same group measured at the

same time every specified interval.

Special operators

Type

Operator Example

Description

Previous value* PRE()

PRE(001)

Obtain the previous measured data of channel 001

Hold**

HOLD():

HOLD(001):TLOG.SUM(002) When the measured value of channel 001 changes from

0 to a nonzero value, maintain the displaying integrated
value of the measured data of channel 002 while the
measured value of channel 001 is nonzero.

Reset**

RESET():

RESET(001):TLOG.SUM(002) When the channel 001 = nonzero, reset the integrated

value of the measured data of channel 002

* Previously measured data or computed data. In the case of computed data, the value is set to 0 when the

computation is reset. At the start of the computation, if the computation was reset, the value is “0”. If it
was not reset, the value is the last value of the previous computation. The value that can be specified
inside the() is limited to the input channel number (001 to 060) or the computation channel number (A01
to A60). Each computing equation can be used once.

** When specifying HOLD(A):B or RESET(A):B, A and B are channel numbers or computing equations.

These can be used once in the beginning of the computing equation.

Limitations in computing equations

Multiple operators can be used in 1 computing equation. But, there are following limitations.
- Number of characters that can be used : 40 characters
- Total number of channel numbers and constants: 16 (Computation erroro ccurs when 16

exceeded, and the computed result becomes +OVER or -OVER)

- Computation channel numbers: Computation channel numbers less than the current

computation channel number can be used as variables within the computing equation.
Example: A02=001+A01

← Computation channel numbers greater than or equal to A03 can

not be used in this computation.

- Statistical operators (TLOG. or CLOG.) can only be used once in 1 computing equation.

Computation channel

A total of 60 computation channels are available (A01 to A60).

Constant

A total of 60 constants can be set (K01 to K60).

Flag (F01 to F16)

Flags can be set in computing equations as constants (1 or 0). Flags are normally 0, but it is set to
1 when a certain event occurs according to the event/action function. For example, if the
computing equation is set to

NOTF01*TLOG.SUM(001),

and FLAG:F01 is set as an edge action of an event/action, F01 becomes 1 when the event occurs.
This causes NOTF01 to become 0 and the SUM of channel 001 to become 0.

Computation range

If the result exceeds

±10

308

during computation, an overflow will occur.

Data applicable for computation

The following data is used for computation.
• Measured data: Specified by channel No. (001 to 560).
• Computed data: Specified by computation channel No. (A01 to A60).
• Constant: Value specified for K01 to K60.
• Group data: Measured data of channels belonging to a group. Specified by group No. (G01 to

G07). This is applicable only for CLOG.

• Communication input data: Data written to the instrument’s memory via communication

interface. Specified by data No. (C01 to C60).

• Data on internal RAM disk: Applicable only for the DR232-1/DR242-1 (equipped with a

floppy disk drive). Measured/computed data saved in the internal RAM disk. Specified by
M001 to M560 (for measured data) or MA01 to MA60 (for computed data).