Yokogawa DR240 User Manual

Page 241

Advertising
background image

13-8

IM DR232-01E

• DC Current Measurement

Apply the 20mA to channel 3

+

Input terminal

(CH3)

DC Current Generator

Short circuit between “+” and “–” terminal in channel 2

+

Input terminal

(CH2)

• Strain measurement

Perform calibrations using 4-gauge method.
Use a 319300 bridge box for the Du500-14’s bridge box.
Wire the strain gauge or the bridge box to the channel 2.

Jumper setting switch

ON

OFF

No.1
No.2

No.3

No.4
No.5

A(+)

B(L)

C(-V)

D(H)

DU500-12/DU500-13

DU500-14

No.1

OFF

No.2

OFF

No.3

OFF

No.4

OFF

No.5

ON

1

2

3

4

5

6

7

8

R4

R2

R3

R1

R2

R4

R3

R1

To R1 to R4, connect resistors with the specifications described below.

Calibration

Resistors R1, R2&R3

Resistor R4

Accuracy

ZERO

120.000

120.000

±0.005%, ±0.3ppm/°C

2k SPAN

120.000

119.521

±0.005%, ±0.3ppm/°C

20k SPAN

120.000

115.294

±0.005%, ±0.3ppm/°C

200k SPAN

120.000

80.000

±0.005%, ±0.3ppm/°C

13.5 Calibration