6 configuring the pulse input channel, Configuring the pulse input channel -13, 01:pulse/rate – Yokogawa DR240 User Manual

Page 116: 01:skip, 01:skip ***set ok, 60:pulse/rate, Span, Filter=on, Set ok

Advertising
background image

5-13

IM DR232-01E

5

Setting the Input Type/Recording Span/Linear Scaling

5.6

Configuring the Pulse Input Channel

D I S P

M O D E

P R I N T

F U N C

I N S

D E L

R A N G E A L A R M

S E T

C H A R T

E N T E R

F E E D

M. F U N C 1

M. F U N C 2

RECORD

ALARM

CHART

KEYLOCK

R E C O R D

001-01:PULSE/RATE

SKIP PULSE

SPAN= 0/ 6000

E S C

• Press the RANGE key to enter the RANGE menu.
• Using

and

, select and/or enter a value for each of the shaded fields shown below.

• To exit any of the following menu items during the procedure, press the MODE (ESC) key.

This returns to the first item of the menu. Note, however, that your new settings and selections
are canceled.

• Reaching the step showing the message “***SET OK***” confirms the latest settings/

selections.

001-01:SKIP

001-01:SKIP

001-01:SKIP

***SET OK***

(configuration complete)

(Main menu item)

(Submenu items)

ENTER

ESC

001-60:PULSE/RATE

001-60:PULSE/RATE

RATE GATE

SPAN=

_ _ _ _ _ _

0/

_ 30000

>Span limit(0 to 30000)

SCL=

_

0/

_ 30000

>SPAN(0/30000)

FILTER=ON

ON OFF

***SET OK***

(configuration complete)

ENTER

ENTER

ENTER

ENTER

ESC

Advertising