Setup=alarm, Alarm rh time=8, Alarm rl time=8 – Yokogawa DR240 User Manual

Page 168: Alarm hys=0.5, Alarm hold=off, A/d intg set*** filter=off, Filter set, Alarm a/d intg filter, Set up

Advertising
background image

10-7

IM DR232-01E

10

Basic Settings (SET UP)

10.3 Select Alarm Interval/Hysteresis/Hold/A/D

Converter Integration Time/Filter

D I S P

M O D E

P R I N T

F U N C

I N S

D E L

R A N G E A L A R M

S E T

C H A R T

E N T E R

F E E D

M. F U N C 1

M. F U N C 2

RECORD

ALARM

CHART

KEYLOCK

R E C O R D

SETUP=ALARM

TVL RECORD ALARM A/D_INTG FILTER RELAY

>Select Setting Parameter

E S C

• Enter the SET UP menu in the following procedure:

1. When the display presents the operation display mode or a menu other than the SET UP

menu, first turn off the power switch.

2. Turn the switch back on while pressing the DISP key and continue pressing the DISP key

for about 5 seconds.

• Select/set

using the

keys.

• If exiting the following lower menu setting halfway, press the MODE (ESC) key. The display

returns to the main menu to which the lower menu belongs. However, newly set or selected
contents are canceled.

• When setting proceeds to the display for the setting end, the newly set or selected contents are

established.

ALARM RH TIME=8

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

ALARM RL TIME=8

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

ALARM HYS=0.5%

0.0 0.1 0.2 0.3 0.4 0.5

0.6 0.7 0.8 0.9 1.0

ALARM HOLD=OFF

OFF ON

***ALARM SET***

S UNIT0:A/D INTG=AUTO

S UNIT0:A/D INTG=AUTO

AUTO 50Hz 60Hz 100ms

***A/D INTG SET***

FILTER=OFF

OFF ON

***FILTER SET***

ESC

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ESC

ESC

ALARM

A/D INTG

FILTER

( M a i n m e n u )

( L o w e r m e n u )

ENTER

SET UP=