Yokogawa DR240 User Manual

Page 240

Advertising
background image

13-7

IM DR232-01E

Trouble-shooting

and Maintenance

13

Required Equipment

Name

Measurement

Accuracy

Recommended

Range

DC Voltage Generator

0V to 50V

0.005% of setting+1

µV

Yokogawa 9000*, 4808

DMM

0V to 50V

0.005% of setting+1

µV

HP3458A

Decade Resistance Box

0.1

Ω to 1000Ω

0.01%

Yokogawa 2793

DC Current Generator

0 to 20mA

0.05% of setting

Yokogawa 7651

* For 0V input, it is necessary to either short the input terminals, or to monitor the output voltage using the

DMM.

Calibrating Conditions

Ambient temperature

: 23

°C±2°C

Ambient humidity

: 55

±10%RH

AC power supply voltage : 100 to 240 VAC (AC power supply model)
DC power supply voltage : 12 to 28 VDC (the subunit of DC power supply model omly)
Power supply frequency : 50/60 Hz

±1%

Warm-up time

: at least 30 minutes for this recorder, and necessary warm-up time for

the used equipment

Connection (terminal screw connections)
• DC Voltage Measurement

• Apply a rated voltage in the voltage range that is to be calibrated to channel 3.

+

DC Voltage Generator

Input terminal

(CH3)

• Short-circuit between “+” and “–” terminals in channel 2.

+

Input terminal

(CH2)

• Temperature Measurement Using RTD

• apply the 100

Ω resistance to channel 5.

B

b

A

Decade Resistance Box

Input terminal

(CH5)

• Short-circuit the resistance input terminals A, B, and b in channel 4.

B

b

A

Input terminal

(CH4)

13.5 Calibration