Chapter 8 setting, displaying and recording alarms, Chapter 8, Setting, displaying and recording alarms – Yokogawa DR240 User Manual

Page 139: 01:1/off

Advertising
background image

8-1

IM DR232-01E

8

Setting, Displaying and Recording Alarms

8.1

Setting Alarms and Relays (including internal
switches)

D I S P

M O D E

P R I N T

F U N C

I N S

D E L

R A N G E A L A R M

S E T

C H A R T

E N T E R

F E E D

M. F U N C 1

M. F U N C 2

RECORD

ALARM

CHART

KEYLOCK

R E C O R D

00101:1/OFF

1=OFF 2=OFF 3=OFF 4=OFF

>

E S C

• Press the [ALARM] key to enter the ALARM menu.
• Select / set

using the

keys.

• To escape from a lower menu, press the MODE (ESC) key. Its main menu will appear,

although new settings / selections will not be kept.

• New settings/selections will be kept when you rach the final display of that setting.

001-01:1/OFF

>Select Channel No.

001-01:1/OFF

>Select Channel No.

001-01:1/OFF

1 2 3 4

001-01:1/OFF

***SET OK***

001-01:1/H_=_ _ _0.00

001-01:1/H_=_ _ _0.00

>Alarm Limit(0.00~100.00)

RELAY=OFF

***SET OK***

RELAY=S01

RELAY=S01

***SET OK***

ENTER

ENTER

ENTER

ESC

ESC

ESC

ENTER

( M a i n m e n u )

( L o w e r m e n u )

Selecting channel Nos.

Channels Nos. A01 to A60 are available for computation channels (optional).

Selecting the alarm item number

Set the alarm item number for each channel as follows:
• Select any number from among 1, 2, 3, and 4.
• Alarm numbers 1 to 4 can be set for one channel.
• For details on setting the alarm printout, see Section 6.5, “Setting the Alarm Printout” on page

6-10.

Note

• If SKIP is selected for the input type or if the computation channels are OFF, alarms cannot be set.

• If the following cases, the alarm is set to OFF:

• If the input type or measuring range in the related channel is changed,

• If computation channels are turned ON/OFF or the computation equation is changed.

• If the recording span for the linear scaling, or linear scaling values is changed, or
• If the standard channel for differential computation between channels is changed.

• Setting the alarm or relay (including internal switches) gives more functions to this recorder. For details,

see Section 9.1, “Setting Event and Action Functions” on page 9-1.