Yokogawa DR240 User Manual

Page 264

Advertising
background image

14-19

IM DR232-01E

Specifications

14

Continuously Applicable Maximum Voltage and Current

Voltage: 250 Vrms
Current: 5 Arms

Crest Factor

3 maximum (600 Vpeak)

Maximum Common Mode Voltage

250 Vrms

Common Mode Voltage Rejection Ratio (Voltage and Current Ranges)

0.02% of span (when 250 V, 45 to 65 Hz is imposed)

Input Resistance

Voltage input: 300 k

Ω minimum for AC voltages

Current input: 1

Ω maximum for AC currents

Filter

Moving average

Wattage Calculation

Use of /M1 computing function

Insulation Resistance

Across output terminals and ground: 20 M

Ω minimum

Withstanding Voltage

Across output terminals and ground: 2,300 V AC (50/60 Hz), 1 min

Normal Operating Temperature/Humidity Ranges

0˚ to 50˚C
20 to 80% RH for an ambient temperature range of 0˚ to 40˚C; 10 to 50% RH for 40˚ to 50˚C
(non-condensing)

Power Consumption

Included in the value of a system's installed main unit or sub-unit

Dimensions

Approximately 57 (W)

× 137 (H) × 99 (D) (mm)

Weight

Approximately 0.5 kg

Installation Category (Overvoltage Category)

CAT II (IEC1010)

14.4 Specifications of Power Monitor Module

Advertising