Wiring Strain Input Signal Lines (Strain Input Module)

Please apply the optional DV450-001 strain conversion cable when using a bridge box or strain
gage without sensor line.

Wiring Diagrams
Model with Built-in Bridge

Model with External Bridge Resistors

CH2

CH3

CH5

CH7

CH9

CH1

CH4

CH6

CH8

CH10

Jumper setup switches

CH1

CH3

CH5

CH7

CH9

CH2

CH4

CH6

CH8

CH10

NDI terminal

Wiring Diagrams
• Single-gauge method

ON

OFF

No.1
No.2

No.3

No.4
No.5

A(+)

B(L)

C(-V)

D(H)

DU500-12/DU500-13
Jumper setup switch

DU500-14

E

e

Rg

R

R

R

Rg

No.1

ON

No.2

ON

No.3

ON

No.4

OFF

No.5

OFF

Rg

R = fixed resistor
r = resistance of leadwire
Rg = resistance of strain gauge
e = output voltage developed across bridge
E = voltage imposed across bridge

1

2

3

4

5

6

7

8

• Single-gauge three-wire method

R

R

R

Rg

r

r

r

E

e

R = fixed resistor
r = resistance of leadwire
Rg = resistance of strain gauge
e = output voltage developed across bridge
E = voltage imposed across bridge