5-10

IM DR232-01E

5.4 Configuring the Measuring Range and

Recording Span or the Linear Scaling of a
Strain Input Channel

D I S P

M O D E

P R I N T

F U N C

I N S

D E L

R A N G E A L A R M

S E T

C H A R T

E N T E R

F E E D

M. F U N C 1

M. F U N C 2

RECORD

ALARM

CHART

KEYLOCK

R E C O R D

001-01:STRAIN/200k

> Select Channel Number

SPAN=–20000/ 20000 * 10

µε

E S C

• Press the RANGE key to enter the RANGE menu.
• Using

and

, select and/or enter a value for each of the shaded fields shown below.

• To exit any of the following menu items during the procedure, press the MODE (ESC) key.

This returns to the first item of the menu. Note, however, that your new settings and selections
are canceled.

• Reaching the step showing the message “***SET OK***” confirms the latest settings/

selections.

Setting the Measuring Range and Recording Span

001-01:STRAIN/200k

>Select Channel No.

001-01:STRAIN/200k

>Select Channel No.

(Main menu item)

001-01:STRAIN/200k

001-01:STRAIN/200k

2k 20k 200k

SPAN=–20000/_20000

>Span limit(–20000 to 20000*10

µε

)

***SET OK***

(configuration complete)

ENTER

ENTER

ESC

Setting Linear Scaling

001-01:STRAIN/200k

>Select Channel No.

001-01:STRAIN/200k

>Select Channel No.

001-01:STRAIN/200k

(Main menu item)

(Submenu items)

001-01:SCL:STRAIN/200k

001-01:SCL:STRAIN/200k

2k 20k 200k

SPAN=–20000/_20000

>Span limit(–20000 to 20000 *10

µε

)

SCL=-100.00/_100.00

***SET OK***

(configuration complete)

ENTER

ESC

ENTER

(Press twice)

ENTER