Scan interval, A/d integration time, Input filter – Yokogawa DR240 User Manual

Page 30

Advertising
background image

2-4

IM DR232-01E

Scan Interval

• The duration of time (one scan) in which the measurement of all channels is carried out, is

called the scan interval.

• This interval can be set to any value from 0.5 second to 60 seconds. The shortest is 300 ch/500

ms (varies with the shortest measurement period of the input module).

A/D Integration Time

This instrument measures the input signal after putting it through an A/D converter. In order to
minimize the noise imposed on the input signal, specific integration times exist.
The integration time can be selected from 20ms (50Hz), 16.7ms (60Hz) and 100ms (10Hz).
When “AUTO” is selected, the integration time will be automatically decided according to the 50/
60Hz frequency of the power supply.
AUTO does not function if the instrument is the subunit (DS400/DS600) of the DC power supply
model (Selecting “AUTO” will set the A/D integration time to 20 ms (50 Hz)). If you are using
the instrument on a 60-Hz power supply, set the A/D integration time to 16.7 ms (60Hz).

Input Filter

A filter can be set ON/OFF to reduce normal mode noise. Effects on normal mode noise are as
follows depending on the filter being ON/OFF (theoretical values).

0dB

–20dB

–40dB

–60dB

–80dB

–100dB

1Hz

10Hz

100Hz

50Hz

300Hz

Slope:–20dB/dec

Slope:–60dB/dec

Filter OFF

Filter ON

Frequency

Attenuation

2.2 Measurement Input Functions