13-9

IM DR232-01E

Trouble-shooting

and Maintenance

13

Entering the calibration mode

1. While pressing and holding the MODE key, turn on the power.

Selecting the subunit and module numbers

2. With

, select the subunit number.

S_UNIT0 : MODULE No=0

>

0 1 2 3 4 5

3. With

, move the cursor.

4. With

, select the module number.

S_UNIT0 : MODULE No=0

0 1 2 3 4 5

5. Press the ENTER key.

Calibrating in CAL/EXEC mode

6. With

, select the CAL/EXEC mode.

ADJUST MODE=CAL/EXEC

>S_UNIT0 : MODULE No=0

DISPLAY CAL/EXEC END

7. Press the ENTER key.
8. With

, select the calibration range.

ADJUST RANGE=20mV

>S_UNIT0 : MODULE No=0

20mV 60mV 200mV 2V 6V 20V 50V Pt : 1mA P

9. Enter the rated value in the selected range.
10.Press the ENTER key to execute calibration. During the calibration, the following display

appears.

A/D Adjusting

>S_UNIT0 : MODULE No=0

20mV 60mV 200mV 2V 6V 20V 50V Pt : 1mA P

11.After calibration is completed, the following display appears.

* * *

CALC END

* * *

>S_UNIT0 : MODULE No=0

>ADJUST RANGE=20mV

12.Press the ENTER key.

13.5 Calibration