Troubleshooting

If an error code appears on the display, see Section 13.4, “Error Codes.”
If servicing is necessary, or if the instrument is not operating correctly though the following
corrective actions have been taken, please contact Yokogawa Engineering Service Corporation.
Addresses may be found on the back cover of this manual.

Problem

Probable Cause

Corrective Action

Recorder does not operate.

Power switch is off.

Turn on the power.

(The recorder carriage and chart do

Power supply is not

Connect properly.

not move.)

connected properly.

Supplied power does not

Use power meeting its requirements.

meet powerrequirements.

Fuse blown.

Replace fuse. (If the fuse blows immediately after turning on the
power, servicing will be required.)

Output beyond its limits

Input specifications are not

Correct input specifications.

Fluctuating indication

correct.

Carriage swung over.

Incorrect measuring range or

Change measuring range or recording span.

recording span

Noise superimposed.

Connect input wiring far away from noise source.

Ground recorder.

Ground measurement object.

Isolate thermocouple from measurement object.

Use shielded wiring for input line.

Change A/D integral times.

Use moving average.

Use input filter.

No countermeasure taken

Install input terminal cover properly.

against ambient temperature

Protect recorder from blowing air of fan.

changes.

Keep temperature changes near input terminals small.

Input connected improperly.

Connect input properly.

Connect module properly.

Tighten screws properly.

Isolate RTD from ground.

Replace disconnected thermocouples.

Recorder connected in parallel

Do not use burnout functions in other instruments.

with other instruments.

Ground recorder and other instruments in the same ground line.

Do not connect recorder in parallel with other instruments
(for use with dual-element TC).

RJC set improperly

Set RJC properly.

(for TC input)

Dot-printing position

Calibrate correctly.

not calibrated correctly.

Other cause.

Contact Yokogawa Engineering Service.

Defective display

Noise superimposed.

Lay input wiring far away from noise source.

Ground recorder.

Ground measurement object.

Isolate TC from measurement object.

Use shielded wires for input line.

Change A/D integration times.

Use input filter.

Data indicates “xxxxxx”

Input module connected

Connect input properly.

improperly

Recorder does not work even with

Key-lock not released.

Disable key-lock functions in setup mode.

operation key pressed.

Recorder set in remote mode.

Set recorder in local mode.

Other cause.

Contact Yokogawa Engineering Service.