14-8

IM DR232-01E

Communication Function

Communication interface is possible by GP-IB, RS-232-C, RS422-A/RS-485, or ethernet.

Power Supply

Rated supply voltage

AC power supply : 100 to 240 VAC (freely selected)
DC power supply : 12 to 28 VDC (only for the subunit (DS400/DS600) of a DC power supply

model)

Operating supply voltage

AC power supply : 90 to 250 VAC
DC power supply : 10 to 32 VDC (only for the subunit (DS400/DS600) of a DC power supply

model)

Rated supply frequency

50/60 Hz

Power consumption

Main unit: max. approx. 130 VA
Sub-unit DS400: max. approx. 55 VA (AC power supply), max. approx. 25 VA (DC power

supply)

Sub-unit DS600: max. approx. 70 VA (AC power supply), max. approx. 25 VA (DC power

supply)

Fuse Ratings

• Main unit

Maximum rated voltage/current: 250 V/2.5 A, Type: Time-lag, Standard: IEC/VDE

• Subunit (The fuse may not be replaced by the customer.)

AC power supply
Maximum rated voltage/current: 250 V/2 A, Type: Time-lag, Standard: IEC/VDE
DC power supply
Maximum rated voltage/current: 250 V/6.3 A, Type: Time-lag, Standard: UL/CSA

Normal Operating Conditions

Supply voltage

AC power supply : 90 to 250 VAC
DC power supply : 10 to 32 VDC (only for the subunit (DS400/DS600) of a DC power supply

model)

Supply frequency

50Hz

±2%, 60Hz ±2%

Ambient temperature

Main unit : 0 to 50

°C

DS400/DS600 Subunit :

when floor-mounted : 0 to 50

°C , when panel-mounted : –10 to 60°C

However, if the subunit DS400/DS600 is a DC power supply model, it can be used in the range
from 0

°C to 50°C regardless of the mounting method.

Ambient humidity

20 to 80%RH for –10 to 40

°C, 10 to 50%RH for 40 to 50°C, 5 to 30%RH for 50 to 60°C (no condensation)

Vibration

10 to 60 Hz, 0.2m/s

2

Shock

Not allowed

Magnetic field

400 A/m max. (50/60Hz)

Position

Unit should be positioned left-right horizontally

Installation location

Room

Installation height

Altitude up to 2,000 m

Installation category based on IEC 1010-1

II*

1

Pollution degree based on IEC 1010-1

2*

2

Warm-up time

At least 30 minutes after power switch-on.

*1 Installation category is the specification of the inpuls withstanding voltage which is also called as

overvoltage category.

*2 Polution degree is the level of foreign body adhesion such as the solid, liquid, and gas which decrease

the wuthstanding voltage, 2 means general indoor atmosphere.

14.1 Specifications of DR232/DR242 and DS400/DS600