3.9 Connecting the Power Cord and Turning the Power ON/OFF

DS400/DS600(when using DC power supply connecter)

This applies only to products with power supply 2 suffix code.

Follow the warnings below to avoid electric shock or damaging the instrument.

WARNING

• Connect the power wires after checking that the power supply is turned

off to prevent electric shock.

• To prevent fire, use wires with cross sectional area of 0.3mm

2

(22AWG)

or more.

CAUTION

• If you connect the + and - terminals in reverse on the DS400/DS600,

the internal fuse will be blown (You cannot replace the fuse by yourself.
The instrument needs servicing in this case). If there is a possibility of
reversing the polarity, insert a fuse (rating shown below) in the wiring.
This will lower the chances of blowing the internal fuse.
Fuse : 250 V/T2.5 A to T4A (time lag), 20 mm glass tube fuse,

I

2

t = 12.5 to 32

(Recommended: A1350EF (250 V/T2.5 A, I

2

t=12.5: SCHURTER:

FST0034.3121))

Connecting procedure

1.Check thst the power switch is turned off.
2.Connect the accessory DC power terminal connector (part No. A1105JC) to the power wire and

the DS400/DS600. Use a power wire with a cross sectional area of 0.3 mm

2

(22AWG) or

more.

21

Soldering

- (0 V DC)

+ (10 to 32 V DC)

DS400/DS600
Rated supply voltage:12 to 28 V DC
Operating supply voltage:10 to 32 V DC
Power consumption:About 25 VA max.

• DS400/DS600: DC power supply model

DC power terminal
connector

DC power
terminal