Pediatric settings – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual
Page 100
Advertising
96
Acuity Central Station
Welch Allyn 1500 Patient Monitor
Pediatric settings
Parameter
Acuity range limit
Acuity lower limit
(default)
Acuity upper limit
(default)
HR
25 - 250
50
150
PR (NIBP)
25 - 250
50
150
PR (IBP)
25 - 250
50
150
PR (SpO
2
)
25 - 250
50
150
SpO
2
SAT (%)
50 - 100
90
100
RR-ECG
2 - 150
10
45
RR-CO
2
2 - 150
10
45
Apnea Delay (s)
6 - 30
N/A
20
etCO
2
(mmhg)
0 - 99
25
60
etCO
2
(kPa)
0 - 13.2
3.0
8.0
inCO
2
(mmhg)
2 - 25
N/A
5
inCO
2
(kPa)
0.2 - 5.0
N/A
0.7
P1 Sys
-30 - 300
75
145
P1 Dia
-30 - 300
35
100
P1 Mean
-30 - 300
50
110
P2 Sys
-30 - 300
15
50
P2 Dia
-30 - 300
5
20
P2 Mean
-30 - 300
10
25
P3 Sys
P3 Dia
P3 Mean
P4 Sys
P4 Dia
P4 Mean
NIBP Sys
30 - 260
75
220
NIBP Dia
20 - 235
35
110
NIBP Mean
20 - 255
50
120
Temp (F)
59 - 113
95.0
100.0
Temp (C)
15 - 45
35.0
37.8
Temp2 (F)
Temp2 (C)
ΔTemp (F)
ΔTemp (C)