ECG alarms and messages

Message

Possible cause

Suggested action

HR – asystole/ASY

No QRS detected for 4 seconds.

Check the patient.
ECG signal lower than 0.5 mV.

HR – ventricular
fibrillation/VF

No organized ventricular rhythm
detected.

Check the patient.
ECG signal lower than 0.5 mV.

HR – artifact

Patient has moved.

Calm the patient.

Bad electrode.

Checking the electrode pads.

Interferences by other devices.

Remove source of the interference.

HR > [upper limit]
HR < [lower limit]

Heart rate higher/lower than
alarm limit.

Check the patient.

HR – lead off

Electrode lose/defective.

Check and reapply/replace electrodes.

Patient cable defective.

Replace the patient cable.