Ecg alarms and messages – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual
Page 53
Advertising
Directions for use
Monitoring and Measurements
49
ECG alarms and messages
Message
Possible cause
Suggested action
HR – asystole/ASY
No QRS detected for 4 seconds.
Check the patient.
ECG signal lower than 0.5 mV.
HR – ventricular
fibrillation/VF
No organized ventricular rhythm
detected.
Check the patient.
ECG signal lower than 0.5 mV.
HR – artifact
Patient has moved.
Calm the patient.
Bad electrode.
Checking the electrode pads.
Interferences by other devices.
Remove source of the interference.
HR > [upper limit]
HR < [lower limit]
Heart rate higher/lower than
alarm limit.
Check the patient.
HR – lead off
Electrode lose/defective.
Check and reapply/replace electrodes.
Patient cable defective.
Replace the patient cable.