134
Technical data
Welch Allyn 1500 Patient Monitor
SpO
2
Nellcor Module
Sensors
Nellcor
®
OxiMax
®
sensors
Amplifier
Fully isolated, defibrillation protected >5kV
Sampling Frequency
62.5 Hz
Display update interval
1 second
Measurement range
SpO
2
1 to 100 %
PR
20 to 250 /min
Accuracy (Probe 70%, to
100 %, 28°C to 42°C)
SpO
2
Adult / pediatric ± 2 digits
Neonate ± 3 digits
PR (no motion)
20 to 250 /min ± 3 digits
Calibration range
70 to 100 % (calibration is fixed, no calibration required)
PR Calculation
Averaged over 4 / 8 / 16 beats