70

Monitoring and Measurements

Welch Allyn 1500 Patient Monitor

SpO

2

settings

The parameter settings are selected with the trim knob, described previously (see

“Settings via a parameter field”

on page 31)

.

The default settings are in bold.

Parameter

Description

SpO

2

alarm pause

When the alarm is paused the message SpO

2

Alarms Paused

appears in the message line. The alarm is paused until the sensor is
reinstalled.

HR/PR source

1

1.

This can also be set in the Setup HR menu.

ECG

,

SpO

2

, P1 (selection of the heart rate source). The source is

displayed in brackets next to the HR parameter.

HR/PR tone

1

Off

, on

2

2.

If SpO

2

is the HR/PR source, the pitch of the beep corresponds to the SpO

2

saturation. A high-pitched beep

indicates a high saturation.

SpO

2

lower/upper limit

Range: 50- 100
Lower limit default 50
Upper limit default 100