Masimo module – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual
Page 139
Directions for use
Technical data
135
Masimo Module
Sensors
Masimo
SET
®
LNCS
®
Amplifier
Fully isolated, defibrillation protected >5kV
Sampling frequency
62.5 Hz
Display update interval
1 second
Signal IQ waveform
A waveform that indicates pulse detection confidence. Values range from
0 to 127 where 0 is low confidence and 127 is high confidence.
Perfusion Index numeric
(PI)
A numeric provided that indicates perfusion. Perfusion is measured in %
and ranges from 0.02% to 20%.
Fast SAT
A mode that enables rapid tracking of arterial oxygen saturation changes
by minimizing the averaging.
Display ranges
Saturation (SpO
2
)
0 to 100%
PR
25 to 240 beats per minute
Averaging Time
2 / 4 / 8 / 10 / 12 / 14 / 16 / 18
Sensitivity
Normal / Max / APOD
Accuracy
SpO
2
, no motion
60 to 80 ± 3%, adults / pediatrics
70 to 100 ± 2%, adults / pediatrics, ± 3% neonates
SpO
2
, motion
70 to 100 ± 3%, adults / pediatrics / neonates
SpO
2
, low perfusion
70 to 100 ± 2%, adults / pediatrics / neonates
PR, no motion
25 to 240 ± 3 bpm, adult s/ pediatrics / neonates
PR, motion
25 to 240 ± 5 bpm, adults / pediatrics / neonates
PR, low perfusion
25 to 240 ± 5 bpm, adults / pediatrics / neonates