Masimo module – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual

Page 139

Advertising
background image

Directions for use

Technical data

135

Masimo Module

Sensors

Masimo

SET

®

LNCS

®

Amplifier

Fully isolated, defibrillation protected >5kV

Sampling frequency

62.5 Hz

Display update interval

1 second

Signal IQ waveform

A waveform that indicates pulse detection confidence. Values range from
0 to 127 where 0 is low confidence and 127 is high confidence.

Perfusion Index numeric
(PI)

A numeric provided that indicates perfusion. Perfusion is measured in %
and ranges from 0.02% to 20%.

Fast SAT

A mode that enables rapid tracking of arterial oxygen saturation changes
by minimizing the averaging.

Display ranges

Saturation (SpO

2

)

0 to 100%

PR

25 to 240 beats per minute

Averaging Time

2 / 4 / 8 / 10 / 12 / 14 / 16 / 18

Sensitivity

Normal / Max / APOD

Accuracy

SpO

2

, no motion

60 to 80 ± 3%, adults / pediatrics
70 to 100 ± 2%, adults / pediatrics, ± 3% neonates

SpO

2

, motion

70 to 100 ± 3%, adults / pediatrics / neonates

SpO

2

, low perfusion

70 to 100 ± 2%, adults / pediatrics / neonates

PR, no motion

25 to 240 ± 3 bpm, adult s/ pediatrics / neonates

PR, motion

25 to 240 ± 5 bpm, adults / pediatrics / neonates

PR, low perfusion

25 to 240 ± 5 bpm, adults / pediatrics / neonates