Technical data, System data, A - technical data – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual

Page 131

Advertising
background image

A

127

Technical data

System data

Manufacturer

SCHILLER AG for Welch Allyn

Monitor name

Welch Allyn

®

1500 Patient Monitor

Dimensions

396 x 284 x 81 mm (15.6 x 11.2 x 3.2 inches)

Weight

5.0 kg (11 lbs) (with lead acid battery)

4.5 kg (9.9 lbs) (with Li-Ion battery)

Mode of operation

Continuous

Power supply

Internal Charger

Voltage

100 – 240 V, 50 – 60 Hz

Power consumption

max 70 VA

Typical battery
operating time

With the battery fully charged, 25°C, display on, NIBP measurement every
15 minutes, and with all parameters ECG/RESP/NIBP/Temp/SpO

2

/IBP(x2)/

CO

2

:

Lead Acid: approximately 1 hour

Li-Ion: approximately 2 hours

Fuses

2 x M 1.6A E 250V

Environmental conditions
for operating

Temperature

10 ºC to 40 ºC (50 ºF to 104 ºF) at relative humidity of 30 to 80 % (non-
condensing)

Atmospheric
pressure

700 to 1060 hPa

Environmental conditions
for transport and storage

Temperature

-10 ºC to 50 ºC (14 ºF to 122 ºF) at relative humidity of 10 to 95 % (non-
condensing)

Atmospheric
pressure

572 to 1060 hPa

Monitor display

Color TFT LCD

Resolution

1024 x 768 pixels

Dimensions

30.7 x 23 cm (12 x 9 ins),15 in diagonal

Speed

6.25/12.5/25 mm/s