Respiration, Temperature, Respiration temperature – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual

Page 136

Advertising
background image

132

Technical data

Welch Allyn 1500 Patient Monitor

Respiration

Temperature

ECG amplifier

Sampling frequency

1000 Hz

Pacemaker detection ± 2 to ± 700 mV / 0.1 to 2 ms

Pacemaker rejection ± 2 to ± 700 mV / 0.1 to 2 ms

Note: Pacemaker signals can differ from one pacemaker to the next. Rate
meters may continue to count the pacemaker rate during occurrences of
cardiac arrest or some arrhythmias, mainly with pacemakers generating
high amplitudes ( > 20 mV) or those generating overshoot. Pacemaker
patients should be kept under close or constant observation.

Protection

Fully isolated, defibrillation protected >5 kV

Line frequency filter

50 or 60 Hz sinusoidal interferences filtered by means of adaptive digital
filtering.

Respiration rate range

0 to 200 breaths / min (pediatric: 0 to 120 breaths per minute)

Connector

Shared with ECG

Signal

28 kHz square wave ± 2.5 V

Patient current

max. 80 :A

Dynamic impedance
range

1 k to1.5 kS, variation of 0.1 to 3 S

Sampling Rate

250 Hz

Respiration rate accuracy ± 1 digit

RR display update interval Max. 2 s

Channels

One or two channels

Sensor

YSI 401, rectal, skin or ear

Amplifier

Fully isolated, defibrillation protected >5kV

Sampling Frequency

125 Hz

Measurement interval

1x per second

Measurement range

15 °C to 45 °C (59 °F to 113°F)

Resolution

0.1 °C (0.1 °F)

Accuracy

+ 0.1° C (+ 0.1° F)