St settings – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual

Page 57

Advertising
background image

Directions for use

Monitoring and Measurements

53

ST settings

The parameter settings are selected with the trim knob, described previously (see

“Settings via a parameter field”

on page 31).

The default settings are in

bold

.

Note

ST measurements can only be displayed when the full arrhythmia and ST option is
licensed, and when ST is enabled in setup parameters (See

“Defining parameter

fields”

on page 25

).

Note

When the monitor is connected to the Acuity Central Station, ST support is
provided by Acuity and the ST option is not displayed on the monitor.

Main menu

Parameter

Description

Setup ST

ST Relearn

Reanalyzes the QRS complexes to determine
base measurement levels.

ST Lead 1

Define the lead for measurement 1 (default lead
II

).

ST lead 2

Define the lead for measurement 2 (default lead
V

).

ST Measurement point

Auto,

60, 80. Define the measurement point (ms

after the j-point).
Note: When Auto is defined, the measurement
point is set to either 60 or 80 ms. This is
dependent on the results of an algorithm that
builds a hysteresis counter based on the
patient's average heart rate - a faster heart rate
is set at 60 ms.

Lower limit / upper limit

Range: -10 to +10
Lower limit default -1
Upper limit default +1

Alarm level

Off, Low, Medium, High.

Print on alarm

Yes, No.

Presets

Select to return all ST settings to the defaults.