Acuity Central Station
97
Neonatal settings
Parameter
Acuity range limit
Acuity lower limit
(default)
Acuity upper limit
(default)
HR
25 - 250
100
200
PR (NIBP)
25 - 250
100
200
PR (IBP)
25 - 250
100
200
PR (SpO
2
)
25 - 250
100
200
SpO
2
SAT (%)
50 - 100
85
95
RR-ECG
3 - 150
10
75
RR-CO
2
3 - 150
10
75
Apnea Delay (s)
6 - 20
N/A
15
etCO
2
(mmhg)
0 - 99
25
60
etCO
2
(kPa)
0 - 13.2
3.0
8.0
inCO
2
(mmhg)
2 - 25
N/A
5
inCO
2
(kPa)
0.2 - 5.0
N/A
0.7
P1 Sys
-30 - 300
50
100
P1 Dia
-30 - 300
30
70
P1 Mean
-30 - 300
35
80
P2 Sys
-30 - 300
15
50
P2 Dia
-30 - 300
5
20
P2 Mean
-30 - 300
10
25
P3 Sys
P3 Dia
P3 Mean
P4 Sys
P4 Dia
P4 Mean
NIBP Sys
25 - 120
50
100
NIBP Dia
15 - 105
30
70
NIBP Mean
15 - 110
35
80
Temp (F)
59 - 113
95.0
100.0
Temp (C)
15 - 45
35.0
37.8
Temp2 (F)
Temp2 (C)
ΔTemp (F)
ΔTemp (C)