Directions for use

Maintenance

101

Battery maintenance

The battery is maintenance free during its normal life.

No maintenance is necessary during normal operation.

If the monitor is not used, check and recharge the batteries every three months. The
battery should not be allowed to fully discharge during storage.

Replace the battery every 2 to 5 years (depending upon application). When the
running time falls substantially under two hours (lithium ion battery), or one hour (lead
acid battery), replace the battery.

Recharging the battery

Totally discharged batteries require the following times to charge:

Lead acid battery: 80% capacity - 2.8 hours, 100% capacity 3.5 hours

Li-Ion battery: 80% capacity - 2.5 hours, 100% capacity - 6.5 hours

It is possible to use the monitor when the battery is being charged; however, the charging
time of the battery will be extended.

1.

Connect the monitor to the mains but do not switch it on.

2. The LEDs for both mains and battery are illuminated.

3. The battery LED is extinguished when the battery is fully charged.

Battery disposal

Batteries must be disposed of in municipally approved areas or sent back to Welch Allyn.
See

“Recycling monitor components”

on page 109.

WARNING

Explosion warning. The battery must not be burned or disposed of in

domestic trash.

WARNING

Flammability and chemical danger. Do not open the battery.

WARNING

Protect the contacts from shorting when disposing of the battery.

Apply non-conducting tape to the contacts.