Etco2 alarms and messages, Etco, Alarms and messages – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual

Page 67

Advertising
background image

Directions for use

Monitoring and Measurements

63

etCO

2

alarms and messages

l

Message

Possible cause

Suggested action

etCO

2

> [upper limit]

etCO

2

< [lower limit]

etCO

2

is higher / lower than

alarm limit.

Check the patient.

RR > [upper limit]
RR < [lower limit]

RR is higher / lower than alarm
limit.

Check the patient.

FiCO

2

> [upper limit]

FiCO

2

has exceeded alarm limit.

Check the patient and ventilator.

CO

2

needs calibration CO

2

measurement reading too

low due to:

Incorrect calibration

Module requires calibration

Carry out calibration.

IPI cannot measure

CO

2

measurement reading too

low.
SpO

2

measurement too low or

absent.

Check the patient.

Check SpO

2

measurement and sensor.