Etco2 alarms and messages, Etco, Alarms and messages – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual
Page 67
Advertising
Directions for use
Monitoring and Measurements
63
etCO
2
alarms and messages
l
Message
Possible cause
Suggested action
etCO
2
> [upper limit]
etCO
2
< [lower limit]
etCO
2
is higher / lower than
alarm limit.
Check the patient.
RR > [upper limit]
RR < [lower limit]
RR is higher / lower than alarm
limit.
Check the patient.
FiCO
2
> [upper limit]
FiCO
2
has exceeded alarm limit.
Check the patient and ventilator.
CO
2
needs calibration CO
2
measurement reading too
low due to:
•
Incorrect calibration
•
Module requires calibration
Carry out calibration.
IPI cannot measure
CO
2
measurement reading too
low.
SpO
2
measurement too low or
absent.
Check the patient.
Check SpO
2
measurement and sensor.