NIBP alarms and messages

Message

Possible cause

Suggested action

NIBP needs service

No NIBP module detected.

Switch off and restart.

Replace monitor.

NIBP artifact
cannot measure

Patient has moved.

Calm the patient.

Max. required pressure is higher
than the initial pressure of
160 mmHg.

1

1.

If the initial pressure is too low, the measurement is immediately restarted and the pressure is increased by
60 mmHg.

Repeat measurement. The monitor will
automatically increase the initial pressure.

Cannot measure
NIBP

Patient has moved.

Check and calm patient.

Very unsteady pulse.

Apply cuff to another extremity with less
movement and steady pulse.

Air tube plugged or leaking.

Check tube and cuff.

NIBP cuff leak

No cuff connected, or cuff or
insufficiently fitted or defective.

Check cuff position.

Check cuff for tightness.

Check if the cuff is connected to the monitor.

NIBP signal low

Blocked tube; kink in the tube.

Check and replace the tube if required.

Cuff not applied correctly.

Reposition/check the cuff.

Pulse too low for good
measurement.

Apply the cuff to another extremity where
the pulse measurement is easier.

NIBP time too long

Inflation time exceeded (max.
135 sec.) due to interferences
because the patient has moved

Check the patient (see also message "cannot
measure").

Repeat the measurement.

NIBPs < [lower limit]

Systolic pressure too low.

Check the patient and alarm limits.

NIBPs > [upper limit]

Systolic pressure too high.

Check the patient and alarm limits.

NIBPd < [lower limit]

Diastolic pressure too low.

Check the patient and alarm limits.

NIBPd > [upper limit]

Diastolic pressure too high.

Check the patient and alarm limits.

NIBPm < [lower limit] Mean pressure too low.

Check the patient and alarm limits.

NIBPm > [upper limit] Mean pressure too high.

Check the patient and alarm limits.