Acuity alarm default settings, Adult settings – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual

Page 99

Advertising
background image

Directions for use

Acuity Central Station

95

Acuity alarm default settings

Adult settings

Parameter

Acuity range limit

Acuity lower limit

(default)

Acuity upper limit

(default)

HR

25 - 250

50

120

PR (NIBP)

25 - 250

50

120

PR (IBP)

25 - 250

50

120

PR (SpO

2

)

25 - 250

50

120

SpO

2

SAT (%)

50 - 100

90

100

RR-ECG

2 - 150

5

30

RR-CO

2

2 - 150

5

30

Apnea Delay (s)

6 - 30

N/A

15

etCO

2

(mmhg)

0 - 99

25

60

etCO

2

(kPa)

0 - 13.2

3.0

8.0

inCO

2

(mmhg)

2 - 25

N/A

5

inCO

2

(kPa)

0.2 - 5.0

N/A

0.7

P1 Sys

-30 - 300

75

220

P1 Dia

-30 - 300

35

110

P1 Mean

-30 - 300

50

120

P2 Sys

-30 - 300

15

50

P2 Dia

-30 - 300

5

20

P2 Mean

-30 - 300

10

25

P3 Sys

P3 Dia

P3 Mean

P4 Sys

P4 Dia

P4 Mean

NIBP Sys

30 - 260

75

220

NIBP Dia

20 - 235

35

110

NIBP Mean

20 - 255

50

120

Temp1 (F)

59 - 113

95.0

100.0

Temp1 (C)

15 - 45

35.0

37.8

Temp2 (F)

Temp2 (C)

ΔTemp (F)

ΔTemp (C)