Acuity alarm default settings, Adult settings – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual
Page 99
Advertising
Directions for use
Acuity Central Station
95
Acuity alarm default settings
Adult settings
Parameter
Acuity range limit
Acuity lower limit
(default)
Acuity upper limit
(default)
HR
25 - 250
50
120
PR (NIBP)
25 - 250
50
120
PR (IBP)
25 - 250
50
120
PR (SpO
2
)
25 - 250
50
120
SpO
2
SAT (%)
50 - 100
90
100
RR-ECG
2 - 150
5
30
RR-CO
2
2 - 150
5
30
Apnea Delay (s)
6 - 30
N/A
15
etCO
2
(mmhg)
0 - 99
25
60
etCO
2
(kPa)
0 - 13.2
3.0
8.0
inCO
2
(mmhg)
2 - 25
N/A
5
inCO
2
(kPa)
0.2 - 5.0
N/A
0.7
P1 Sys
-30 - 300
75
220
P1 Dia
-30 - 300
35
110
P1 Mean
-30 - 300
50
120
P2 Sys
-30 - 300
15
50
P2 Dia
-30 - 300
5
20
P2 Mean
-30 - 300
10
25
P3 Sys
P3 Dia
P3 Mean
P4 Sys
P4 Dia
P4 Mean
NIBP Sys
30 - 260
75
220
NIBP Dia
20 - 235
35
110
NIBP Mean
20 - 255
50
120
Temp1 (F)
59 - 113
95.0
100.0
Temp1 (C)
15 - 45
35.0
37.8
Temp2 (F)
Temp2 (C)
ΔTemp (F)
ΔTemp (C)