Temperature monitoring

Depending on the sensor type, the sensor can be applied to the ear, the skin or the
rectum.

To achieve a reliable measured value, independent of the measuring site, the
measurement duration must be at least 2 minutes.

Temperature settings

The parameter settings are selected with the trim knob, described previously (see

“Settings via a parameter field”

on page 31)

.

The default settings are in bold.

Temperature alarm and messages

Parameter

Description

TEMP Display Mode

T1 and T2,

T1 only, or T1 and

ΔT

Temp Units

°C

or °F

T1 and T2 lower / upper limit

Range: 15°C to 45°C (59°F and 113°F)
Lower limit default: 35°C (95°F)
Upper limit default: 37.8°C (100°F)

ΔT lower / upper limit

Range: 15°C to 45°C (59°F and 113°F)
Lower limit default: 0.0°C (0.0°F)
Upper limit default: 0.1°C (0.2°F)

Note

Alarms can only be set for the temperature parameters that are currently on
display. So for example, if T1 and delta T were selected for display, there would be
no T2 alarm settings.

Message

Possible cause

Suggested action

TEMP unplugged

TEMP not connected to the
monitor

Connect sensor.

TEMP needs service

The monitor has detected an
error

Switch monitor Off/On or replace monitor.

TEMP out of range

The temperature is outside the
measuring range of the monitor.

Check the patient or alarm limits.

Sensor or monitor problem

Check the sensor and monitor. Switch the
monitor off/on.

TEMP < [lower limit]

Temperature too low

Check the patient and alarm limits.

TEMP > [upper limit]

Temperature too high

Check the patient and alarm limits.