Measured values

ECG

Patient cable

3-lead, 5-lead, 12-lead cable

Automatic 3, 5 and 12 lead detection

Lead fault detection

AAMI 6 pin and 12 pin connectors

Leads

Simultaneous, synchronous recording of up to nine active electrodes
giving 12 leads

Filters

Mains

50 Hz / 60 Hz / off

Bandwidth

0.05 Hz / 0.5 Hz , 35 Hz / 150 Hz

Input impedance

> 2.58 MS

Heart rate range

15 to 300 beats/min

QRS tone

On / Off

Protection

ESU and defibrillator protected

Lead display

Selectable leads
Selection of 1 to 5 simultaneous leads

Display update interval

1 second

Lead fail sense current

< 0.5 :A

Tall T-wave rejection

max. amplitude of the T-wave according to IEC 60601-2-27 chapter
50.102.17: 4 mV

HR averaging method

The average of the last 16 beats is used, when RR interval corresponds to a
HR of < 48 bpm.

The average of the last 4 beats is used, when RR interval corresponds to a
HR of $48 bpm.

HR accuracy

± 5 % or ± 5 bpm (whichever is greater)

HR meter response time

Change from 80 to 120 bpm: 11s

Change from 80 to 40 bpm: 11s

Response to Irregular
rhythm

A1: 80/min
A2: 60/min
A3: 120/min
A4: 90/min
(according to IEC specification 60601-2-27, 6.8.2.bb)

Time to Alarm for
tachycardia

B1 and B2: 3 s
(according to IEC specification 60601-2-27, 6.8.2.bb)

Sensitivity

According to ANSI/AAMI EC13 / IEC60601-2-27