84

Monitoring and Measurements

Welch Allyn 1500 Patient Monitor

4. Inject the solution. The temperature waveform is displayed and the CO value given in

measurement 1.

5. Repeat to obtain 3- 5 measurements

.

Editing the measurements

The average value is displayed in the bottom middle of the screen and is updated after
every measurement or if a measurement is deleted.

After five measurements, when another measurement is taken, the latest
measurement takes the place of measurement 5 and all measurements are shifted
left (and measurement 1 is deleted).

Any measurement can be deleted by selecting and pressing the trim knob. If another
measurement is taken, the later measurements are shifted left as above and the
measurement deleted is replaced by the one higher.

Select Review curves to view the temperature waveforms of all measurements.

Note

Follow you facility’s guidelines for injecting the solution.

Temperature waveform

CO Measurements 1 to 5

Averaged value of all
measurements

View all temperature
curves