Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual

Page 62

Advertising
background image

58

Monitoring and Measurements

Welch Allyn 1500 Patient Monitor

WARNING

If too much moisture enters the sampling line (i.e., from ambient

humidity or breathing of unusually humid air), and the sampling line cannot be
cleared, the message Blockage appears in the message area. Replace the
sampling line once the sampling line blockage message appears.

Caution

In high-altitude environments, etCO

2

values may be lower than values

observed at sea level. When using the monitor in high altitude environments, it is
advisable to consider adjusting EtCO

2

alarm settings accordingly.

Caution

Microstream

®

etCO

2

sampling lines are designed for single patient

use, and are not to be reprocessed. Do not attempt to clean, disinfect, sterilize or
flush any part of the sampling line as this can cause damage to the monitor.

Caution

Dispose of sampling lines according to standard operating procedures

or local regulations for the disposal of contaminated medical waste.

Caution

Before use, carefully read the Microstream etCO

2

sampling lines

Directions for Use.

Caution

Only use Microstream etCO

2

sampling lines to ensure the monitor

functions properly.

Caution

Dispose of Microstream etCO

2

sampling lines according to standard

operating procedures or local regulations for the disposal of contaminated
medical waste.

Caution

During nebulization or suction for Intubated patients, in order to avoid

moisture buildup and sampling line occlusion, remove the sampling line luer
connector from the monitor.

Caution

Replace the sampling line according to hospital protocol or when a

blockage is indicated by the monitor. Excessive patient secretions or a build-up of
liquids in the airway tubing may occlude the sampling line, requiring more
frequent replacement.

Advertising