Respiration rate, Respiration rate settings – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual
Page 59

Directions for use
Monitoring and Measurements
55
Respiration rate
If the RR should be measured via the ECG instead of etCO
2
, the etCO
2
measurement
field must be deactivated as follows:
Press the Setup button
, select Parameters and deactivate etCO
2
(No).
Respiration rate settings
The parameter settings are selected with the trim knob, described previously (see
“Settings via a parameter field”
.
The default settings are in bold.
Note
The RR measurement field is not displayed if the etCO
2
field is enabled.
Note
The RR signal is measured via the R (RA) and F (LL) electrodes of the ECG cable
(impedance measurement). After the patient is connected, about 30 seconds can
elapse before a reliable value is displayed.
Main menu
Parameter
Description
RESP Enabled
Display RR measurement
Yes, No. When the respiration is not
enabled, (Disabled) is displayed in the
measurement screen.
Apnea
Apnea time
6, 10, 15, 20, 25, 30 seconds.
Setup RR
Lower limit / upper limit
Range: 2- 150
Lower limit default 5
Upper limit default 30