Arrhythmia, Arrhythmia settings – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual
Page 54

50
Monitoring and Measurements
Welch Allyn 1500 Patient Monitor
Arrhythmia
Arrhythmia settings
1.
Press the Setup button
.
2. Select Arrhythmia.
The default settings are in bold.
Note
The full arrhythmia menu is only displayed when the full arrhythmia option is
licensed. When the full arrhythmia option has not been enabled, settings are
made for VFib, Asystol and Vtach only.
Note
The arrhythmia menu is not available when the monitor is connected to Acuity.
Main menu
Parameter
Description
Setup Arrhythmia
VFib
Alarm Level - Lethal (cannot be changed), Print
on alarm Yes/No.
Asystole
Alarm Level - Lethal (cannot be changed), Print
on alarm Yes/No.
VTach
VTach Limit 100 to 200 (125), Alarm Level - Lethal
(cannot be changed), Print on alarm Yes/No.
PVC run
PVC run 3 to 6 (6) Alarm Level (High, Medium,
Low, Off), Print on alarm Yes/No.
VRyhthm
Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.
Couplet
Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.
PVC/min
PVC/min 1to 30 (30) Alarm Level (High, Medium,
Low, Off), Print on alarm Yes/No.
Bigeminy
Alarm Level (High, Medium, Low, Off), Print on
alarm Yes/No.