Monitoring and Measurements

Welch Allyn 1500 Patient Monitor

ECG menu settings

The default settings are in bold.

Note

The parameter settings are selected with the trim knob. (See

“Settings via a

parameter field”

on page 31

).

Main menu

Parameter

Description

Setup HR

ECG Lead

Lead selection I,

II

, III, V, AVL, AVR, AVF

Size

0.5, 1, 2, 4 mV/cm

HR/PR source

1

1.

This can also be set in the setup SpO

2

menu.

ECG

, SpO

2

, P1

HR/PR Tone

1

Off

/ on

2

2.

If SpO

2

is selected, the pitch of the beep corresponds to the SpO

2

saturation. A high pitched beep indicates a high

saturation.

ECG Filter

Select Diagnostic or Monitor
Two predefined filter settings can be selected. These defined
filter options define the cut off frequency for the Myogram,
Baseline and Mains filters. The filter definitions and the
corresponding filter cut-offs are defined as follows:

Baseline

Myogram

Mains

Diagnostic

0.05 Hz

150 Hz

as set

Monitor

0.50

35 Hz

as set

Single ECG

Yes/No, select lead

I

, II, III, or V. Select this option to analyze one

ECG lead.

Pacer Display

Yes/No displays pacer pulses relative to time but not
representative of either amplitude, duration or polarity.

Analyzer Pacer

Yes/No - not applicable.

HR lower / upper
limit

3

3.

The lower limit cannot be set to a higher value than the value set for the upper limit, and vice versa.

Range: 25 - 250
Lower limit default 50
Upper limit default 120

Cal

Generates a simulated 1 mV calibration impulse on the curve.

Note

A setting is available in the administrator menu that prevents the HR / PR alarm
(upper / lower limit) from being switched off (see

“Administrator” on page 118

).