Maintenance
103
Cleaning instructions and cleaning solutions
Equipment
Cleaning instructions
Approved cleaning solutions
Monitor
1
1.
The equipment can be disinfected to comply with OSHA requirements for cleaning and decontaminating spills of blood
and other body fluids. (Federal OSHA blood borne pathogens standard: 29 CFR 1910.1030, 12/6/91.)
Wipe with a nearly
dry cloth moistened with
cleaning solution.
Thoroughly wipe off any excess cleaning
solution. Do not let cleaning solution run into
connector openings or crevices.
2
2.
If liquid gets into the connectors, dry the area with warm air and then verify all monitoring functions.
70 % solution isopropyl alcohol;
neutral mild detergent solution;
all products designed for cleaning
plastic.
ECG cable, extension
cable
Consult manufacturer’s instructions.
Mild detergent solution; also
consult manufacturer’s
instructions.
SpO
2
cable, extension
cable
Consult manufacturer’s instructions.
Consult manufacturer’s
instructions.
Other accessories
Consult manufacturer’s instructions.
Consult manufacturer’s
instructions.