Alarms
Display of alarms
During initial powerup
No alarms are displayed if no patient is being monitored.
During monitoring
There are three alarm priorities:
Alarm type
Priority
LED visual alarm
indicator
Audible signal
Display
Technical
Low
Blue
Single deep tone
every 2 seconds
Text display in the alarm status field at the
bottom.
Parameter
Medium
Yellow (flashes with
parameter field)
Two tone high/low
every second.
Text display in the alarm status field at the
bottom.
Yellow flashing parameter field.
Parameter
High
Red (flashes with
parameter field)
Three high tones
every second.
Text display in the alarm status field at the bottom
Red flashing parameter field.
Lethal
High
Red (flashes with
parameter field)
Three high tones
every second.
Text display in the alarm status field at the bottom
Red flashing parameter field.
Visual alarm indicator
Text Display
Parameter Field