Spo2 messages, Messages – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual

Page 76

Advertising
background image

72

Monitoring and Measurements

Welch Allyn 1500 Patient Monitor

SpO

2

messages

Message

Possible cause

Suggested action

SpO

2

– check sensor Defective SpO

2

sensor

Replace the sensor.

Incorrect settings in the monitor Check the monitor settings.

SpO

2

– check sensor

placement

Poor sensor contact or the
sensor has fallen off

Check the contact between the sensor and
the patient.

Sensor is disturbed by ambient
light

Cover the sensor.

Sensor defective (red light on
the sensor is not lit)

Replace the sensor.

SpO

2

Low Perfusion

1

1.

When the monitor has a Masimo module, the message SpO2 low perfusion only occurs when the perfusion index
feature is enabled.

Sensor not properly applied

Check the sensor and reapply.

Fingernail varnish on the finger

Remove fingernail varnish.

Thick skin

Change finger.

Sensor failed

Change sensor.

Weak Signal

see SpO

2

Low Perfusion above

see SpO

2

Low Perfusion above.

SpO

2

unplugged

SpO

2

not connected to the

monitor

Connect sensor.

SpO

2

artifact

Patient has moved

Calm the patient.

Hemodynamic interference

Apply sensor to another extremity.

Too thin skin

Apply sensor to a larger finger.

SpO

2

< [lower limit]

SpO

2

too low

Check the patient and alarm limits.

SpO

2

> [upper limit]

SpO

2

too high

Check the patient and alarm limits.

SpO2 low signal IQ

2

2.

The SpO2 low signal IQ message is only given when the monitor has a Masimo module and only occurs if Signal IQ
waveform feature is enabled.

Poor quality signal

Check the sensor and reapply. Move sensor
to obtain a better quality signal.

PR < [lower limit]

Pulse rate too low

Check the patient and alarm limits.

PR > [upper limit]

Pulse rate too high

Check the patient and alarm limits.