Ibp alarm defaults, Match the label (see – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual
Page 79
Advertising
Directions for use
Monitoring and Measurements
75
IBP alarm defaults
Label / Parameter
Adult
Lower Limit
Adult
Upper Limit
Pediatric
Lower Limit
Pediatric
Upper Limit
Neonate
Lower Limit
Neonate
Upper Limit
P1, P2, P3, P4
SYS
P1, P2, P3, P4
DIA
P1, P2, P3, P4
Mean
6
14
2
10
2
10
-4
6
-4
2
-4
2
0
10
0
4
0
4
ART SYS
ART
DIA
ART Mean
75
220
75
145
50
100
35
110
35
100
30
70
50
120
50
110
35
80
CVP SYS
CVP DIA
CVP Mean
6
14
2
10
2
10
-4
6
-4
2
-4
2
0
10
0
4
0
4
PA
SYS
PA
DIA
PA
Mean
10 34
10
34
10 34
0
16
0
16
0
16
0
20
0
20
0
20
ICP
SYS
ICP
DIA
ICP
Mean
6
14
2
10
2
10
-4
6
-4
2
-4
2
0
10
0
4
0
4
RA
SYS
RA
DIA
RA
Mean
6
14
2
10
2
10
-4
6
-4
2
-4
2
0
10
0
4
0
4
LA
SYS
LA
DIA
LA
Mean
6
14
2
10
2
10
-4
6
-4
2
-4
2
0
10
0
4
0
4
UA
SYS
UA
DIA
UA
Mean
50
100
30
70
35
80
UV
SYS
UV
DIA
UV
Mean
2
10
-4
2
0
4