IBP alarm defaults

Label / Parameter

Adult
Lower Limit

Adult
Upper Limit

Pediatric
Lower Limit

Pediatric
Upper Limit

Neonate
Lower Limit

Neonate
Upper Limit

P1, P2, P3, P4

SYS

P1, P2, P3, P4

DIA

P1, P2, P3, P4

Mean

6

14

2

10

2

10

-4

6

-4

2

-4

2

0

10

0

4

0

4

ART SYS
ART

DIA

ART Mean

75

220

75

145

50

100

35

110

35

100

30

70

50

120

50

110

35

80

CVP SYS
CVP DIA
CVP Mean

6

14

2

10

2

10

-4

6

-4

2

-4

2

0

10

0

4

0

4

PA

SYS

PA

DIA

PA

Mean

10 34

10

34

10 34

0

16

0

16

0

16

0

20

0

20

0

20

ICP

SYS

ICP

DIA

ICP

Mean

6

14

2

10

2

10

-4

6

-4

2

-4

2

0

10

0

4

0

4

RA

SYS

RA

DIA

RA

Mean

6

14

2

10

2

10

-4

6

-4

2

-4

2

0

10

0

4

0

4

LA

SYS

LA

DIA

LA

Mean

6

14

2

10

2

10

-4

6

-4

2

-4

2

0

10

0

4

0

4

UA

SYS

UA

DIA

UA

Mean

50

100

30

70

35

80

UV

SYS

UV

DIA

UV

Mean

2

10

-4

2

0

4