Replacing the fuses, Fuse types, Changing the fuse – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual

Page 110: Fuse types changing the fuse

Advertising
background image

106

Maintenance

Welch Allyn 1500 Patient Monitor

Replacing the fuses

Fuse types

Changing the fuse

1.

Disconnect the monitor from the mains.

2. Release the fuse holder by gently squeezing the side retaining clips and remove the

fuse holder.

3. Replace both fuses. Re-insert the fuse holder until the two side clips snap in place.

WARNING

Disconnect the monitor from the mains before changing the fuses.

WARNING

Blown fuses must only be replaced with the fuse types indicated in

the below table.

Voltage range

Number

Fuse type

WA Part No.

Manufacturer Part No.

100-240 VAC

2

M 1.6A E 250V

4.210004

Schurter Inc, FSM 0034.2518

M= Medium time lag
E= Enhanced breaking capacity