Replacing the fuses, Fuse types, Changing the fuse – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual
Page 110: Fuse types changing the fuse
Advertising
106
Maintenance
Welch Allyn 1500 Patient Monitor
Replacing the fuses
Fuse types
Changing the fuse
1.
Disconnect the monitor from the mains.
2. Release the fuse holder by gently squeezing the side retaining clips and remove the
fuse holder.
3. Replace both fuses. Re-insert the fuse holder until the two side clips snap in place.
WARNING
Disconnect the monitor from the mains before changing the fuses.
WARNING
Blown fuses must only be replaced with the fuse types indicated in
the below table.
Voltage range
Number
Fuse type
WA Part No.
Manufacturer Part No.
100-240 VAC
2
M 1.6A E 250V
4.210004
Schurter Inc, FSM 0034.2518
M= Medium time lag
E= Enhanced breaking capacity