Troubleshooting, General – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual
Page 111
Advertising
Directions for use
Maintenance
107
Troubleshooting
General
Alarm/Condition
Possible cause
Suggested action
Recorder out of paper Paper tray empty
Insert new paper
Check paper
Paper jammed
Check paper
Recorder needs
service
Printer error; paper not
transported correctly; wrong
paper
Check printer; check paper; wrong paper;
paper not inserted correctly; have printer
replaced.
Battery low
Battery capacity too low
Connect to the mains and recharge
battery.
No HR/PR tone
HR/PR tone source setting
Set tone source to on.