Safety – Welch Allyn 1500 Patient Monitor v.1.4.X - User Manual User Manual

Page 83

Advertising
background image

Directions for use

Monitoring and Measurements

79

Safety

WARNING

Carefully read the manufacturer's catheter instructions before

carrying out measurements.

WARNING

When an invasive catheter is introduced into an arterial vessel, the

circulation in the terminal vessels must be checked at regular intervals.

WARNING

When applying the kit to the patient, make sure that no air

penetrates the system.

WARNING

PAC use is complicated by arrhythmias, infection, pulmonary artery

rupture, and right heart valve damage.

WARNING

Single-use accessories must not be reused.

WARNING

It is important that the CO Computation Constant (see

“Cardiac

output settings”

on page 81) of the catheter is entered correctly to ensure

accurate measurements for the specific Edwards-compatible catheter used.

Caution

Only use Edwards-compatible catheters

Caution

Only use Baxter/Edwards/Abbott "in-line" injectate sensors

Caution

It is important that the injectate temperature is in the range specified

for the computation constant.

Caution

It is important that the injectate volume is the same as that specified in

the computation constant.

Note

The kit and operating procedure vary according to manufacturer. Please consult
the manufacturer's documentation for connection.

Note

For warm-up time/ready for measurement and displacement for invasive
transducers, refer to the documentation of the transducer manufacturer.