Trend data

The monitor records up to 24 hours of trend data at one-minute intervals. The trend values
are displayed at the user defined interval (see below) and additionally after every manual
NIBP measurement.

Trend data is deleted when a new patient is entered.

When the memory is full, the oldest trend data is overwritten.

The display interval for the trend table can selected for intervals of one minute, five
minutes, 15 minutes, one hour, or four hours.

Trend data display and settings

Press the Trend button

.

Previous measurements are displayed using the up and down icons.

Use the trim knob to select the trend display interval with the pull down menu in the
Interval

setting.

The Clear option deletes all stored trend data.

The Print option prints the trend data.

Note

The data recorded will depend on licensed and enabled options and the number
of IBP and temperature connectors installed on the monitor (see

“Defining

parameter fields”

on page 25).