Taking a single NIBP measurement

1.

Press the NIBP start/stop button

.

2. The measurement can be stopped at any time by pressing the button again.

Automatic blood pressure measurement

1.

Press the NIBP measurement interval button

.

2. Select the interval between 3 minutes and 60 minutes, and confirm your selection

with OK.

3. The message NIBP interval – xx minutes is displayed.

4. The first measurement is started by pressing the NIBP start/stop button

.

Note

After exiting the standby mode, ensure that the NIBP intervals are re-armed by
manually starting an NIBP measurement.

Note

These settings are reset when the monitor is switched off and automatic
measurement must again be defined when the monitor is switched on.

Note

The monitor sets the maximum pressure as follows:

Adults: 270 mmHg.

Pediatric: 180 mmHg.

Neonate: 150 mmHg.

The patient mode is defined in the setup menu (

“Patient mode”

on page 117).