Preparing the Oridion sensor

Basic principles for choosing microstream

CO

2

sampling lines

When choosing Microstream CO

2

sampling lines, the following should be considered

:

The condition of the patient (ventilated or not ventilated)

If the patient is ventilated, whether ventilation is humidified or non-humidified

Patient’s size and weight

The probability that the patient will switch between oral and nasal breathing

Duration of use

For best results, for short term monitoring, use Microstream CO

2

sampling lines with

orange connectors. For long term monitoring, use Microstream CO

2

sampling lines

with yellow connectors.” Products that include an “H” in the name are intended for
long term use.

Select the appropriate sampling line and connect it to the monitor before putting on the
patient. Be sure to follow the directions for provided with the sampling line.

For further information, please contact your local representative.

Connecting a sampling line

The appropriate sampling line must be connected to the monitor and to the patient.
Connect as follows:

1.

Slide open the sampling line input connector shutter and
connect the appropriate sampling line.

2. Screw the sampling line connector into the monitor clockwise

until it can no longer be turned.

3. Connect the sampling line to the patient as described in the

Directions for Use supplied with the sampling line.

When the sampling line is connected, the monitor will immediately begin to search for
breaths, but it will not indicate a No Breath condition before any valid breaths have
occurred.

CO

2

data displayed by the monitor

The monitor Home screen displays real time CO

2

data. The displayed data includes:

Real time etCO

2

values along with selected units and alarm settings

Respiration rate (RR) in breaths per minute and alarm settings

Real-time FiCO

2

values along with selected units

and alarm settings

CO

2

Waveform if enabled (see

“Waveform display”

on page 28)

IPI value if enabled (see following).