Section 11. Troubleshooting

ADV6600

Y S I

Environmental

Page 143


SYMPTOM


POSSIBLE CAUSE


ACTION


REF

Probe not properly calibrated

Follow DO cal procedures

Membrane not properly
installed or may be punctured

Follow 6562 setup procedure

DO probe electrodes require
cleaning

Follow DO cleaning procedure.
Use 6035 maintenance kit

Water in probe connector

Dry connector; reinstall probe

Algae or other contaminant
clinging to DO probe

Rinse DO probe with clean water

Barometric pressure entry is
incorrect

Repeat DO cal procedure

Cal at extreme temperature

Recalibrate at (or near) sample
temperature

DO Charge too high ( >100)
1. Anodes polarized (tarnished)
2. Probe left on continuously

Enable DO charge parameter in
the Sonde report menu. Run
sonde, if charge is over 100,
recondition probe with 6035
Maintenance Kit. Follow DO
cleaning procedure.

DO Charge too low (<25)
Insufficient electrolyte.

Replace electrolyte and
membrane.

DO probe has been damaged

Replace 6562 probe

Dissolved Oxygen
reading unstable
or inaccurate

Internal failure

Return sonde for service

Probe requires cleaning,

Follow probe cleaning procedure

Probe requires calibration

Follow cal procedures

pH probe reference junction has
dried out from improper
storage.

Soak probe in tap water or buffer
until readings become stable

Water in probe connector

Dry connector; reinstall probe

Probe has been damaged

Replace probe

Calibration solutions out of
spec or contaminated with
other solution

Use new calibration solutions

ORP fails Zobell check

Take into account temperature
dependence of Zobell solution
readings

pH, ORP,
Chloride,
Ammonium, or
Nitrate readings
are unstable or
inaccurate. Error
messages appear
during calibration.

Internal failure

Return sonde for service