Section 4. Calibration and Diagnostics

ADV6600

Y S I

Environmental

Page 49


With the ADV6600 in an upright position:

1. Ensure that a gasket is installed in the gasket groove of the calibration cup bottom cap and that

the bottom cap is securely tightened. Do not over-tighten as this could cause damage to the
threaded portions of the bottom cap and tube
.

4-5.5. Recommended Volumes of Calibration Reagents


The approximate volumes of the reagents are specified below for both inverted and upright
orientations.


Probe to Calibrate

Inverted

Upright

Conductivity 425ml

N/A

Temperature 425ml

N/A

pH/ORP 425ml

N/A

ISE 450ml

N/A

Turbidity, Chlorophyll, Rhodamine WT

N/A

675ml


4-6. Water Quality Sensors - Calibration Procedures


1. Connect the ADV6600 to

a computer.

2. Launch the ADVantage

6600 Software for
Windows.

3. Choose Connect to

System. Verify the COM
setting and Baud Rate.
Click on Connect.

4. Select Change System

Settings under the Data
Collection
screen menu.
A Quick Setup dialog box will appear.

5. Click on Sonde Menu. A window will appear displaying the Main Menu of the Water Quality

firmware.