Section 3. Installation and Setup

ADV6600

Y S I

Environmental

Page 26


The number of items on this menu depends on the sensors that are available and enabled on your
sonde. Below we describe every possible item on this menu. Your sonde may not
have every item described.

NOTE: The user is cautioned to retain the default settings unless advised to change them by YSI
Technical Support. Changes should only be necessary for special applications.

TDS constant
Default: 0.65

This selection allows you to set the constant used to calculate TDS. TDS
in g/L is calculated by multiplying this constant times the specific
conductance in mS/cm. This item will only appear if the conductivity
sensor is enabled in the “Sensors enabled” menu.

Salinity
Default: 0

This selection allows you to input a manually-acquired value of salinity
for calculating DO mg/L. This item is not used or displayed if the
conductivity sensor is enabled in the Sensors menu.

Pres psi
Default: 0

This selection allows you to set a value of pressure for calculating other
parameters like salinity.

DO temp co %/C
Default: 1.1

This selection allows a user to input the dissolved oxygen temperature
coefficient. Do not change this value unless you consult YSI Technical
Support. This item will only appear if the DO sensor is enabled in the
Sensors menu.

DO warm up sec
Default: 60

This selection allows a user to change the warm up period for the
dissolved oxygen sensor. This setting should always be equal to the
averaging interval established in the Quick Setup dialog box.

Wipes
Default: 1

This selection will determine the number of cleaning cycles that will
occur when an optical probe wiper is activated manually or