Section 3. Installation and Setup

ADV6600

Y S I

Environmental

Page 25

Parameter Default Operating

Range

Comments

NO

3

A

2.543

Not checked

Cl J

99.5

Not checked

Cl S

-0.195

-0.217 to -0.15

Cl A

2.543

Not checked

Turb Offset

0

-10 to 10

Turb A1

500

Turb M1

500

0.6 to 1.5

Range is ratio of M1 to A1

Turb A2

1000

Turb M2

1000

0.6 to 1.5

Range is ratio of (M2-M1) to
(A2-A1)

Chl Offset

0

-30 to 20

Chl A1

500

Chl M1

500

0.6 to 1.5

Range is ratio of M1 to A1

Chl A2

1000

Chl M2

1000

0.6 to 1.5

Range is ratio of (M2-M1) to
(A2-A1)

Rhod Offset

0

-10 to 10

Rhod A1

500

Rhod M1

500

0.6 to 1.5

Range is ratio of M1 to A1

Rhod A2

1000

Rhod M2

1000

0.6 to 1.5

Range is ratio of (M2-M1) to
(A2-A1)



From the Advanced menu, select 2-Setup to activate or deactivate Auto sleep RS-232. Activation
of this feature enables a power savings system within the sonde. When enabled, power is only
applied to the sensors during sampling or calibration. The most important aspect of this feature is
its effect on the dissolved oxygen protocol. For this reason, the feature should be activated for most
applications. The selection should only be deactivated if you are running spot sampling studies
where the Averaging Interval and Sample Interval are set to the same value (see later).

From the Advanced menu, select 3-Sensor to display and change user-configurable constants as
shown. Select the appropriate number to view these parameters.