11. Select Change System Settings. A Quick Setup dialog box will appear.
12. Click on Sonde Menu. A window will appear displaying the Main Menu of the 6600 firmware.
13. To activate the parameter, it is necessary to exit the menu structure of the sonde software to the
command line prompt (“#”). The “#” sign appears after you press Esc from the Main menu and
then answer Y to exit the menu.
14. Once the “#” prompt is shown on the display, type dolocal 1 and press Enter. The screen will
show “OK” to indicate that the action was successful and return another “#” sign.
15. Type menu at the second “#” sign and press Enter to return to the sonde software menu
structure.
After the above action, the parameter “DOsat %Local” will be present in your Report menu. If you
want to remove the parameter, return to the “#” prompt, type dolocal 0, and press